Method Comparison and Bias Estimation between Chemiluminescence Immunoassay and Latex-Enhanced Immunoturbidimetric Assay for Detecting FDP and D-Dimer

LI Yuqiong, PENG Yiyi, XU Yang, CHEN Jiajin, TANG Xudong, REN Biqiong

Journal of Hunan Normal University(Medical Science) ›› 2026, Vol. 23 ›› Issue (3) : 127-132.

PDF(2669 KB)
PDF(2669 KB)
Journal of Hunan Normal University(Medical Science) ›› 2026, Vol. 23 ›› Issue (3) : 127-132.
Laboratory Medicine

Method Comparison and Bias Estimation between Chemiluminescence Immunoassay and Latex-Enhanced Immunoturbidimetric Assay for Detecting FDP and D-Dimer

  • LI Yuqiong1, PENG Yiyi1, XU Yang1, CHEN Jiajin1, TANG Xudong2, REN Biqiong1,3
Author information +
History +

Abstract

Objective Using the CLSI EP09c guideline as the standard, this study evaluated the comparability and bias of measurement procedures for detecting fibrin/fibrinogen degradation products (FDP) and D-dimers (DD) in plasma using chemiluminescence and latex immunoturbidimetry methods, providing a reference basis for mutual recognition of FDP and DD results obtained using different methods. Methods Latex immunoturbidimetry was used as the reference method (X) and chemiluminescence immunoassay as the test method (Y). Passing-Bablok regression was used to perform linear fitting of the results, and the bias at the selected concentration levels was calculated separately. Using 1/2 TEa (the total allowable error permitted by the National Health Commission Clinical Laboratory Center's interlaboratory quality assessment) as the evaluation criterion, the test results were subjected to methodological comparison and bias assessment, and the correlation, consistency, and superiority of the two measurement procedures were assessed. Results The bias of chemiluminescence method was -9.59% (<17.5%) for FDP at 5.0 mg/L and 14.29% (<25.0%) for DD at 0.5 mg/L, both within acceptable ranges. The chi-square test results of DD and FDP showed that the two methods had a good correlation (P<0.001), The Kappa test results of FDP were k=0.66(P<0.001) and the Kappa test of DD was k=0.85(P<0.001), both indicating good agreement, and the McNemar test showed no significant difference between the two methodologies (P>0.05), and the two measurement procedures were comparable. Conclusion The measurement procedures for FDP and DD using the two methods show good consistency and correlation. The bias at the upper limit of the clinical reference range is within an acceptable range, meeting clinical testing requirements, and mutual recognition is possible within a certain range.

Key words

EP09c / fibrin degradation products / D-Dimer / chemiluminescence method / latex immunoturbidimetry / measurement procedure

Cite this article

Download Citations
LI Yuqiong, PENG Yiyi, XU Yang, CHEN Jiajin, TANG Xudong, REN Biqiong. Method Comparison and Bias Estimation between Chemiluminescence Immunoassay and Latex-Enhanced Immunoturbidimetric Assay for Detecting FDP and D-Dimer[J]. Journal of Hunan Normal University(Medical Science). 2026, 23(3): 127-132

References

[1] FRANCHINI M, FOCOSI D, PEZZO MP, et al.How we manage a high D-dimer[J]. Haematologica, 2024, 109(4): 1035-1045.
[2] CHEN Y, ZHANG S, ZENG Q, et al.Plasma fibrinogen, D-dimer, and fibrin degradation product as biomarkers of rheumatoid arthritis[J]. Sci Rep, 2021, 11: 16846.
[3] ZHU B, ZHANG L, DU W, et al.Levels of fibrin degradation products at admission with acute ischemic stroke correlate with the NIH stroke scale score 1 h after intravenous thrombolysis[J]. Front Neurol, 2021, 12: 651867.
[4] 王梦园, 刘家敏. D-二聚体与纤维蛋白原对急性脑卒中患者静脉溶栓预后的预测价值[J]. 现代医学与健康研究电子杂志, 2025, 9(17): 100-103.
[5] HORIUCHI H, MORISHITA E, URANO T, et al.COVID-19-related thrombosis in Japan: final report of a questionnaire-based survey in 2020[J]. J Atheroscler Thromb, 2021, 28(4): 406-416.
[6] DONG W, TANG Y, LEI M, et al.The effect of perioperative sequential application of multiple doses of tranexamic acid on postoperative blood loss after PLIF: a prospective randomized controlled trial[J]. Int J Surg, 2024, 110(4): 2122-2133.
[7] 欧元祝, 陈宝荣, 居漪. 临床化学检测的标准化现状[J]. 检验医学, 2021, 36(3): 240-244.
[8] CLSI. Measurement procedure comparison and bias estimation using patient samples[S]. 3rd ed. Wayne, PA: Clinical and Laboratory Standards Institute, 2018.
[9] BEVAN S, LONGSTAFF C.Is it possible to make a common reference standard for D-dimer measurements? Communication from the ISTH SSC subcommittee on fibrinolysis[J]. J Thromb Haemost, 2022, 20(2): 498-507.
[10] WAUTHIER L, FAVRESSE J, HARDY M, et al.D-dimer testing: A narrative review[J]. Adv Clin Chem, 2023, 114: 151-223.
[11] FOLSOM AR, GOTTESMAN RF, APPIAH D, et al.Plasma d-dimer and incident ischemic stroke and coronary heart disease: The Atherosclerosis Risk in Communities Study[J]. Stroke, 2016, 47(1): 18-23.
[12] SIMES J, ROBLEDO KP, WHITE HD, et al.D-dimer predicts long-term cause-specific mortality, cardiovascular events, and cancer in patients with stable coronary heart disease: LIPID Study[J]. Circulation, 2018, 138(7): 712-723.
[13] DI NISIO M, VAN ES N, BÜLLER H R. Deep vein thrombosis and pulmonary embolism[J]. Lancet, 2016, 388(10063): 3060-3073.
[14] JOHNSON ED, SCHELL JC, RODGERS GM.The D-dimer assay[J]. Am J Hematol, 2019, 94(7): 833-839.
[15] FREUND Y, CHAUVIN A, JIMENEZ S, et al.Effect of a diagnostic strategy using an elevated and age-adjusted D-dimer threshold on thromboembolic events in emergency department patients with suspected pulmonary embolism: A randomized clinical trial[J]. JAMA, 2021, 326(21): 2141-2149.
[16] VAN ES N, TAKADA T, KRAAIJPOEL N, et al.Diagnostic management of acute pulmonary embolism: a prediction model based on a patient data meta-analysis[J]. Eur Heart J, 2023, 44(32): 3073-3081.
[17] SUZUKI T, DISTANTE A, ZIZZA A, et al.Diagnosis of acute aortic dissection by D-dimer: the international registry of acute aortic dissection substudy on biomarkers (IRAD-Bio) experience[J]. Circulation, 2009, 119(20): 2702-2707.
[18] SODECK G, DOMANOVITS H, SCHILLINGER M, et al.D-dimer in ruling out acute aortic dissection: a systematic review and prospective cohort study[J]. Eur Heart J, 2007, 28(24): 3067-3075.
[19] 王欢, 徐令清, 朱振杰. 乳胶增强免疫比浊法与电化学发光法检测降钙素原的对比分析[J]. 吉林医学, 2018, 39(8): 1410-1413.
[20] 中国研究型医院学会血栓与止血专业委员会. D-二聚体实验室检测与临床应用中国专家共识[J]. 中华医学杂志, 2023, 103(35): 2743-2756.
PDF(2669 KB)

Accesses

Citation

Detail

Sections
Recommended

/

〈 〉